EasyVoid: A Handheld Coandă Effect Uroflowmeter for Accurate and Portable Measurement of Urinary Flow Dynamics

Chang, Ray, et al. "EasyVoid: A Handheld Coandă Effect Uroflowmeter for Accurate and Portable Measurement of Urinary Flow Dynamics." IEEE Journal of Translational Engineering in Health and Medicine 14 (2026): 210-224.

DOI: https://doi.org/10.1109/JTEHM.2026.3679932

Abstract

Objective: Diagnosis of lower urinary tract symptoms (LUTS) often relies on subjective questionnaires, conventional uroflowmetry or invasive urodynamic study, each with inherent limitations.Methods and procedures: This work presents EasyVoid, a novel handheld device that employs the Coandă effect to enable non-invasive and quantitative measurement of urinary flow. EasyVoid uses a passive fluidic mechanism in which the urinary stream adheres to a curved surface and transfers angular momentum to a helical rotor, allowing direct estimation of flow rate and volume without pressure sensors or weighing systems.Results: EasyVoid simultaneously records key voiding parameters, including flow rate, flow pattern, voiding time, and voided volume, with high temporal precision. In controlled bench-top trials using simulated urine events, measurements obtained from EasyVoid showed strong agreement with standard uroflowmetry across multiple flow patterns.Conclusion: These results demonstrate the potential of EasyVoid for portable and quantitative assessment of urinary flow dynamics, offering a promising solution for remote urinary health monitoring.

Previous
Previous

Fluid-Squid: DIY Multiplexed Imaging of Cells and Tissues

Next
Next

Ballistic Microscopy (BaM)